Definisi Alat-Alat Kesehatan (Neurology, FDA)

Group: Neurology, FDA Version

Subpart B–Neurological Diagnostic Devices

Sec. 882.1020 Rigidity analyzer.

(a)Identification. A rigidity analyzer is a device for quantifying the extent of the rigidity of a patient’s limb to determine the effectiveness of drugs or other treatments.

(b)Classification. Class II (performance standards).

Sec. 882.1030 Ataxiagraph.

(a)Identification. An ataxiagraph is a device used to determine the extent of ataxia (failure of muscular coordination) by measuring the amount of swaying of the body when the patient is standing erect and with eyes closed.

(b)Classification. Class I (general controls).

[44 FR 51730-51778, Sept. 4, 1979, as amended at 66 FR 46952, Sept. 10, 2001]

Sec. 882.1200 Two-point discriminator.

(a)Identification. A two-point discriminator is a device with points used for testing a patient’s touch discrimination.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to 882.9. The device is also exempt from the current good manufacturing practice requirements of the quality system regulation in part 820 of this chapter, with the exception of 820.180 of this chapter, with respect to general requirements concerning records, and 820.198 of this chapter, with respect to complaint files.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 54 FR 25051, June 12, 1989; 65 FR 2319, Jan. 14, 2000]

Sec. 882.1240 Echoencephalograph.

(a)Identification. An echoencephalograph is an ultrasonic scanning device (including A-scan, B-scan, and doppler systems) that uses noninvasive transducers for measuring intracranial interfaces and blood flow velocity to and in the head.

(b)Classification. Class II (performance standards).

Sec. 882.1275 Electroconductive media.

(a)Identification. Electroconductive media are the conductive creams or gels used with external electrodes to reduce the impedance (resistance to alternating current) of the contact between the electrode surface and the skin.

(b)Classification. Class II (performance standards).

Sec. 882.1310 Cortical electrode.

(a)Identification. A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain’s electrical activity.

(b)Classification. Class II (performance standards).

Sec. 882.1320 Cutaneous electrode.

(a)Identification. A cutaneous electrode is an electrode that is applied directly to a patient’s skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

(b)Classification. Class II (performance standards).

Sec. 882.1330 Depth electrode.

(a)Identification. A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.

(b)Classification. Class II (performance standards).

Sec. 882.1340 Nasopharyngeal electrode.

(a)Identification. A nasopharyngeal electrode is an electrode which is temporarily placed in the nasopharyngeal region for the purpose of recording electrical activity.

(b)Classification. Class II (performance standards).

Sec. 882.1350 Needle electrode.

(a)Identification. A needle electrode is a device which is placed subcutaneously to stimulate or to record electrical signals.

(b)Classification. Class II (performance standards).

Sec. 882.1400 Electroencephalograph.

(a)Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient’s brain obtained by placing two or more electrodes on the head.

(b)Classification. Class II (performance standards).

Sec. 882.1410 Electroencephalograph electrode/lead tester.

(a)Identification. An electroencephalograph electrode/lead tester is a device used for testing the impedance (resistance to alternating current) of the electrode and lead system of an electroencephalograph to assure that an adequate contact is made between the electrode and the skin.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 61 FR 1123, Jan. 16, 1996; 66 FR 38807, July 25, 2001]

Sec. 882.1420 Electroencephalogram (EEG) signal spectrum analyzer.

(a)Identification. An electroencephalogram (EEG) signal spectrum analyzer is a device used to display the frequency content or power spectral density of the electroencephalogram (EEG) signal.

(b)Classification. Class I (general controls).

[44 FR 51730-51778, Sept. 4, 1979, as amended at 66 FR 46953, Sept. 10, 2001]

Sec. 882.1430 Electroencephalograph test signal generator.

(a)Identification. An electroencephalograph test signal generator is a device used to test or calibrate an electroencephalograph.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63011, Dec. 7, 1994; 66 FR 38807, July 25, 2001]

Sec. 882.1460 Nystagmograph.

(a)Identification. A nystagmograph is a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball.

(b)Classification. Class II (performance standards).

Sec. 882.1480 Neurological endoscope.

(a)Identification. A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.

(b)Classification. Class II (performance standards).

Sec. 882.1500 Esthesiometer.

(a)Identification. An esthesiometer is a mechanical device which usually consists of a single rod or fiber which is held in the fingers of the physician or other examiner and which is used to determine whether a patient has tactile sensitivity.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to 882.9. The device is also exempt from the current good manufacturing practice requirements of the quality system regulation in part 820 of this chapter, with the exception of 820.180 of this chapter, with respect to general requirements concerning records, and 820.198 of this chapter, with respect to complaint files.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 54 FR 25051, June 12, 1989; 65 FR 2319, Jan. 14, 2000]

Sec. 882.1525 Tuning fork.

(a)Identification. A tuning fork is a mechanical device which resonates at a given frequency and is used to diagnose hearing disorders and to test for vibratory sense.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9. The device is also exempt from the current good manufacturing practice requirements of the quality system regulation in part 820 of this chapter, of this chapter, with the exception of 820.180, with respect to general requirements concerning records, and 820.198, with respect to complaint files.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 54 FR 25051, June 12, 1989; 66 FR 38807, July 25, 2001]

Sec. 882.1540 Galvanic skin response measurement device.

(a)Identification. A galvanic skin response measurement device is a device used to determine autonomic responses as psychological indicators by measuring the electrical resistance of the skin and the tissue path between two electrodes applied to the skin.

(b)Classification. Class II (performance standards).

Sec. 882.1550 Nerve conduction velocity measurement device.

(a)Identification. A nerve conduction velocity measurement device is a device which measures nerve conduction time by applying a stimulus, usually to a patient’s peripheral nerve. This device includes the stimulator and the electronic processing equipment for measuring and displaying the nerve conduction time.

(b)Classification. Class II (performance standards).

Sec. 882.1560 Skin potential measurement device.

(a)Identification. A skin potential measurement device is a general diagnostic device used to measure skin voltage by means of surface skin electrodes.

(b)Classification. Class II (performance standards).

Sec. 882.1570 Powered direct-contact temperature measurement device.

(a)Identification. A powered direct-contact temperature measurement device is a device which contains a power source and is used to measure differences in temperature between two points on the body.

(b)Classification. Class II (performance standards).

Sec. 882.1610 Alpha monitor.

(a)Identification. An alpha monitor is a device with electrodes that are placed on a patient’s scalp to monitor that portion of the electroencephalogram which is referred to as the alpha wave.

(b)Classification. Class II (performance standards).

Sec. 882.1620 Intracranial pressure monitoring device.

(a)Identification. An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.

(b)Classification. Class II (performance standards).

Sec. 882.1700 Percussor.

(a)Identification. A percussor is a small hammerlike device used by a physician to provide light blows to a body part. A percussor is used as a diagnostic aid during physical examinations.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9. The device is also exempt from the current good manufacturing practice requirements of the quality system regulation in part 820 of this chapter, with the exception of 820.180, with respect to general requirements concerning records, and 820.198, with respect to complaint files.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 54 FR 25051, June 12, 1989; 59 FR 63011, Dec. 7, 1994; 66 FR 38807, July 25, 2001]

Sec. 882.1750 Pinwheel.

(a)Identification. A pinwheel is a device with sharp points on a rotating wheel used for testing pain sensation.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 54 FR 25051, June 12, 1989; 65 FR 2319, Jan. 14, 2000]

Sec. 882.1790 Ocular plethysmograph.

(a)Identification. An ocular plethysmograph is a device used to measure or detect volume changes in the eye produced by pulsations of the artery, to diagnose carotid artery occlusive disease (restrictions on blood flow in the carotid artery).

(b)Classification. Class III (premarket approval).

(c) Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before September 21, 2004, for any ocular plethysmograph that was in commercial distribution before May 28, 1976. Any other ocular plethysmograph shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 52 FR 17739, May 11, 1987; 69 FR 34920, June 23, 2004]

Sec. 882.1825 Rheoencephalograph.

(a)Identification. A rheoencephalograph is a device used to estimate a patient’s cerebral circulation (blood flow in the brain) by electrical impedance methods with direct electrical connections to the scalp or neck area.

(b)Classification. Class III (premarket approval).

(c)Date PMA or notice of completion of a PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before December 26, 1996 for any rheoencephalograph that was in commercial distribution before May 28, 1976, or that has, on or before December 26, 1996 been found to be substantially equivalent to a rheoencephalograph that was in commercial distribution before May 28, 1976. Any other rheoencephalograph shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 52 FR 17740, May 11, 1987; 61 FR 50708, Sept. 27, 1996]

Sec. 882.1835 Physiological signal amplifier.

(a)Identification. A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).

(b)Classification. Class II (performance standards).

Sec. 882.1845 Physiological signal conditioner.

(a)Identification. A physiological signal conditioner is a device such as an integrator or differentiator used to modify physiological signals for recording and processing.

(b)Classification. Class II (performance standards).

Sec. 882.1855 Electroencephalogram (EEG) telemetry system.

(a)Identification. An electroencephalogram (EEG) telemetry system consists of transmitters, receivers, and other components used for remotely monitoring or measuring EEG signals by means of radio or telephone transmission systems.

(b)Classification. Class II (performance standards).

Sec. 882.1870 Evoked response electrical stimulator.

(a)Identification. An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.

(b)Classification. Class II (performance standards).

Sec. 882.1880 Evoked response mechanical stimulator.

(a)Identification. An evoked response mechanical stimulator is a device used to produce a mechanical stimulus or a series of mechanical stimuli for the purpose of measuring a patient’s evoked response.

(b)Classification. Class II (performance standards).

Sec. 882.1890 Evoked response photic stimulator.

(a)Identification. An evoked response photic stimulator is a device used to generate and display a shifting pattern or to apply a brief light stimulus to a patient’s eye for use in evoked response measurements or for electroencephalogram (EEG) activation.

(b)Classification. Class II (performance standards).

Sec. 882.1900 Evoked response auditory stimulator.

(a)Identification. An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.

(b)Classification. Class II (performance standards).

Sec. 882.1925 Ultrasonic scanner calibration test block.

(a)Identification. An ultrasonic scanner calibration test block is a block of material with known properties used to calibrate ultrasonic scanning devices (e.g., the echoencephalograph).

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63011, Dec. 7, 1994; 66 FR 38807, July 25, 2001]

Sec. 882.1950 Tremor transducer.

(a)Identification. A tremor transducer is a device used to measure the degree of tremor caused by certain diseases.

(b)Classification. Class II (performance standards).

Subparts C-D [Reserved]

Subpart E–Neurological Surgical Devices

Sec. 882.4030 Skull plate anvil.

(a)Identification. A skull plate anvil is a device used to form alterable skull plates in the proper shape to fit the curvature of a patient’s skull.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63011, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Sec. 882.4060 Ventricular cannula.

(a)Identification. A ventricular cannula is a device used to puncture the ventricles of the brain for aspiration or for injection. This device is frequently referred to as a ventricular needle.

(b)Classification. Class I (general controls). When made only of surgical grade stainless steel, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 65 FR 2319, Jan. 14, 2000]

Sec. 882.4100 Ventricular catheter.

(a)Identification. A ventricular catheter is a device used to gain access to the cavities of the brain for injection of material into, or removal of material from, the brain.

(b)Classification. Class II (performance standards).

Sec. 882.4125 Neurosurgical chair.

(a)Identification. A neurosurgical chair is an operating room chair used to position and support a patient during neurosurgery.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63012, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Sec. 882.4150 Scalp clip.

(a)Identification. A scalp clip is a plastic or metal clip used to stop bleeding during surgery on the scalp.

(b)Classification. Class II (performance standards).

Sec. 882.4175 Aneurysm clip applier.

(a)Identification. An aneurysm clip applier is a device used by the surgeon for holding and applying intracranial aneurysm clips.

(b)Classification. Class II (performance standards).

Sec. 882.4190 Clip forming/cutting instrument.

(a)Identification. A clip forming/cutting instrument is a device used by the physician to make tissue clips from wire stock.

(b)Classification. Class I. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63012, Dec. 7, 1994]

Sec. 882.4200 Clip removal instrument.

(a)Identification. A clip removal instrument is a device used to remove surgical clips from the patient.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63012, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Sec. 882.4215 Clip rack.

(a)Identification. A clip rack is a device used to hold or store surgical clips during surgery.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 54 FR 25051, June 12, 1989; 59 FR 63012, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Sec. 882.4250 Cryogenic surgical device.

(a)Identification. A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.

(b)Classification. Class II (performance standards).

Sec. 882.4275 Dowel cutting instrument.

(a)Identification. A dowel cutting instrument is a device used to cut dowels of bone for bone grafting.

(b)Classification. Class II (performance standards).

Sec. 882.4300 Manual cranial drills, burrs, trephines, and their accessories

(a)Identification. Manual cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments that are used without a power source on a patient’s skull.

(b)Classification. Class II (performance standards).

Sec. 882.4305 Powered compound cranial drills, burrs, trephines, and their accessories.

(a)Identification. Powered compound cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient’s skull. The instruments employ a clutch mechanism to disengage the tip of the instrument after penetrating the skull to prevent plunging of the tip into the brain.

(b)Classification. Class II (performance standards).

Sec. 882.4310 Powered simple cranial drills, burrs, trephines, and their accessories.

(a)Identification. Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient’s skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.

(b)Classification. Class II (performance standards).

Sec. 882.4325 Cranial drill handpiece (brace).

(a)Identification. A cranial drill handpiece (brace) is a hand holder, which is used without a power source, for drills, burrs, trephines, or other cutting tools that are used on a patient’s skull.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 61 FR 1123, Jan. 16, 1996; 66 FR 38808, July 25, 2001]

Sec. 882.4360 Electric cranial drill motor.

(a)Identification. An electric cranial drill motor is an electrically operated power source used with removable rotating surgical cutting tools or drill bits on a patient’s skull.

(b)Classification. Class II (performance standards).

Sec. 882.4370 Pneumatic cranial drill motor.

(a)Identification. A pneumatic cranial drill motor is a pneumatically operated power source used with removable rotating surgical cutting tools or drill bits on a patient’s skull.

(b)Classification. Class II (performance standards).

Sec. 882.4400 Radiofrequency lesion generator.

(a)Identification. A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.

(b)Classification. Class II (performance standards).

Sec. 882.4440 Neurosurgical headrests.

(a)Identification. A neurosurgical headrest is a device used to support the patient’s head during a surgical procedure.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63012, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Sec. 882.4460 Neurosurgical head holder (skull clamp).

(a)Identification. A neurosurgical head holder (skull clamp) is a device used to clamp the patient’s skull to hold head and neck in a particular position during surgical procedures.

(b)Classification. Class II (performance standards).

Sec. 882.4500 Cranioplasty material forming instrument.

(a)Identification. A cranioplasty material forming instrument is a roller used in the preparation and forming of cranioplasty (skull repair) materials.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63012, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Sec. 882.4525 Microsurgical instrument.

(a)Identification. A microsurgical instrument is a nonpowered surgical instrument used in neurological microsurgery procedures.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63012, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Sec. 882.4535 Non-powered neurosurgical instrument.

(a)Identification. A nonpowered neurosurgical instrument is a hand instrument or an accessory to a hand instrument used during neurosurgical procedures to cut, hold, or manipulate tissue. It includes specialized chisels, osteotomes, curettes, dissectors, elevators, forceps, gouges, hooks, surgical knives, rasps, scissors, separators, spatulas, spoons, blades, blade holders, blade breakers, probes, etc.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63012, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Sec. 882.4545 Shunt system implantation instrument.

(a)Identification. A shunt system implantation instrument is an instrument used in the implantation of cerebrospinal fluid shunts, and includes tunneling instruments for passing shunt components under the skin.

(b)Classification. Class I (general controls). When made only of surgical grade stainless steel, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 65 FR 2319, Jan. 14, 2000]

Sec. 882.4560 Stereotaxic instrument.

(a)Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient’s brain, spinal cord, or other part of the nervous system.

(b)Classification. Class II (performance standards).

Sec. 882.4600 Leukotome.

(a)Identification. A leukotome is a device used to cut sections out of the brain.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63012, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Sec. 882.4650 Neurosurgical suture needle.

(a)Identification. A neurosurgical suture needle is a needle used in suturing during neurosurgical procedures or in the repair of nervous tissue.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 54 FR 25051, June 12, 198965 FR 2319, Jan. 14, 2000]

Sec. 882.4700 Neurosurgical paddie.

(a) A neurosurgical paddie is a pad used during surgery to protect nervous tissue, absorb fluids, or stop bleeding.

(b)Classification. Class II (performance standards).

[44 FR 51730-51778, Sept. 4, 1979, as amended at 69 FR 10332, Mar. 5, 2004]

Sec. 882.4725 Radiofrequency lesion probe.

(a)Identification. A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.

(b)Classification. Class II (performance standards).

Sec. 882.4750 Skull punch.

(a)Identification. A skull punch is a device used to punch holes through a patient’s skull to allow fixation of cranioplasty plates or bone flaps by wire or other means.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to 882.9. This exemption does not apply to powered compound cranial drills, burrs, trephines, and their accessories classified under 882.4305.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 65 FR 2319, Jan. 14, 2000]

Sec. 882.4800 Self-retaining retractor for neurosurgery.

(a)Identification. A self-retaining retractor for neurosurgery is a self-locking device used to hold the edges of a wound open during neurosurgery.

(b)Classification. Class II (performance standards).

Sec. 882.4840 Manual rongeur.

(a)Identification. A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.

(b)Classification. Class II (performance standards).

Sec. 882.4845 Powered rongeur.

(a)Identification. A powered rongeur is a powered instrument used for cutting or biting bone during surgery involving the skull or spinal column.

(b)Classification. Class II (performance standards).

Sec. 882.4900 Skullplate screwdriver.

(a)Identification. A skullplate screwdriver is a tool used by the surgeon to fasten cranioplasty plates or skullplates to a patient’s skull by screws.

(b)Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 59 FR 63012, Dec. 7, 1994; 66 FR 38808, July 25, 2001]

Subpart F–Neurological Therapeutic Devices

Sec. 882.5030 Methyl methacrylate for aneurysmorrhaphy.

(a)Identification. Methyl methacrylate for aneurysmorrhaphy (repair of aneurysms, which are balloonlike sacs formed on blood vessels) is a self-curing acrylic used to encase and reinforce intracranial aneurysms that are not amenable to conservative management, removal, or obliteration by aneurysm clip.

(b)Classification. Class II (performance standards).

Sec. 882.5050 Biofeedback device.

(a)Identification. A biofeedback device is an instrument that provides a visual or auditory signal corresponding to the status of one or more of a patient’s physiological parameters (e.g., brain alpha wave activity, muscle activity, skin temperature, etc.) so that the patient can control voluntarily these physiological parameters.

(b)Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter when it is a prescription battery powered device that is indicated for relaxation training and muscle reeducation and prescription use, subject to 882.9.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 63 FR 59229, Nov. 3, 1998]

Sec. 882.5070 Bite block.

(a)Identification. A bite block is a device inserted into a patient’s mouth to protect the tongue and teeth while the patient is having convulsions.

(b)Classification. Class II (performance standards).

Sec. 882.5150 Intravascular occluding catheter.

(a)Identification. An intravascular occluding catheter is a catheter with an inflatable or detachable balloon tip that is used to block a blood vessel to treat malformations, e.g., aneurysms (balloonlike sacs formed on blood vessels) of intracranial blood vessels.

(b)Classification. Class III (premarket approval).

(c)Date PMA or notice of completion of a PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before December 26, 1996 for any intravascular occluding catheter that was in commercial distribution before May 28, 1976, or that has, on or before December 26, 1996 been found to be substantially equivalent to an intravascular occluding catheter that was in commercial distribution before May 28, 1976. Any other intravascular occluding catheter shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 52 FR 17740, May 11, 1987; 61 FR 50708, Sept. 27, 1996]

Sec. 882.5175 Carotid artery clamp.

(a)Identification. A carotid artery clamp is a device that is surgically placed around a patient’s carotid artery (the principal artery in the neck that supplies blood to the brain) and has a removable adjusting mechanism that protrudes through the skin of the patient’s neck. The clamp is used to occlude the patient’s carotid artery to treat intracranial aneurysms (balloonlike sacs formed on blood vessels) or other intracranial vascular malformations that are difficult to attach directly by reducing the blood pressure and blood flow to the aneurysm or malformation.

(b)Classification. Class II (performance standards).

Sec. 882.5200 Aneurysm clip.

(a)Identification. An aneurysm clip is a device used to occlude an intracranial aneurysm (a balloonlike sac formed on a blood vessel) to prevent it from bleeding or bursting.

(b)Classification. Class II (performance standards).

Sec. 882.5225 Implanted malleable clip.

(a)Identification. An implanted malleable clip is a bent wire or staple that is forcibly closed with a special instrument to occlude an intracranial blood vessel or aneurysm (a balloonlike sac formed on a blood vessel), stop bleeding, or hold tissue or a mechanical device in place in a patient.

(b)Classification. Class II (performance standards).

Sec. 882.5235 Aversive conditioning device.

(a)Identification. An aversive conditioning device is an instrument used to administer an electrical shock or other noxious stimulus to a patient to modify undesirable behavioral characteristics.

(b)Classification. Class II (performance standards).

Sec. 882.5250 Burr hole cover.

(a)Identification. A burr hole cover is a plastic or metal device used to cover or plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery.

(b)Classification. Class II (performance standards).

Sec. 882.5275 Nerve cuff.

(a)Identification. A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).

(b)Classification. Class II (performance standards).

Sec. 882.5300 Methyl methacrylate for cranioplasty.

(a)Identification. Methyl methacrylate for cranioplasty (skull repair) is a self-curing acrylic that a surgeon uses to repair a skull defect in a patient. At the time of surgery, the surgeon initiates polymerization of the material and forms it into a plate or other appropriate shape to repair the defect.

(b)Classification. Class II (performance standards).

Sec. 882.5320 Preformed alterable cranioplasty plate.

(a)Identification. A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.

(b)Classification. Class II (performance standards).

Sec. 882.5330 Preformed nonalterable cranioplasty plate.

(a)Identification. A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.

(b)Classification. Class II (performance standards).

Sec. 882.5360 Cranioplasty plate fastener.

(a)Identification. A cranioplasty plate fastener is a screw, wire, or other article made of tantalum, vitallium, or stainless steel used to secure a plate to the patient’s skull to repair a skull defect.

(b)Classification. Class II (performance standards).

Sec. 882.5500 Lesion temperature monitor.

(a)Identification. A lesion temperature monitor is a device used to monitor the tissue temperature at the site where a lesion (tissue destruction) is to be made when a surgeon uses a radiofrequency (RF) lesion generator and probe.

(b)Classification. Class II (performance standards).

Sec. 882.5550 Central nervous system fluid shunt and components.

(a)Identification. A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

(b)Classification. Class II (performance standards).

Sec. 882.5800 Cranial electrotherapy stimulator.

(a)Identification . A cranial electrotherapy stimulator is a device that applies electrical current to a patient’s head to treat insomnia, depression, or anxiety.

(b)Classification. Class III (premarket approval).

(c)Date a PMA or notice of completion of a PDP is required. No effective date has been established of the requirement for premarket approval. See 882.3.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 52 FR 17740, May 11, 1987; 60 FR 43969, Aug. 24, 1995; 62 FR 30457, June 4, 1997; 73 FR 34860, June 19, 2008]

Sec. 882.5810 External functional neuromuscular stimulator.

(a)Identification. An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient’s gait.

(b)Classification. Class II (performance standards).

Sec. 882.5820 Implanted cerebellar stimulator.

(a)Identification. An implanted cerebellar stimulator is a device used to stimulate electrically a patient’s cerebellar cortex for the treatment of intractable epilepsy, spasticity, and some movement disorders. The stimulator consists of an implanted receiver with electrodes that are placed on the patient’s cerebellum and an external transmitter for transmitting the stimulating pulses across the patient’s skin to the implanted receiver.

(b)Classification. Class III (premarket approval).

(c)Date premarket approval application (PMA) or notice of completion of a product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before September 26, 1984. Any implanted cerebellar stimulator that was not in commercial distribution before May 28, 1976, or that has not on or before September 26, 1984 been found by FDA to be substantially equivalent to an implanted cerebellar stimulator that was in commercial distribution before May 28, 1976 shall have an approved PMA or declared completed PDP in effect before beginning commercial distribution.

[44 FR 51730-51778, Sept. 4, 1979 and 49 FR 26574, June 28, 1984]

Sec. 882.5830 Implanted diaphragmatic/phrenic nerve stimulator.

(a)Identification. An implanted diaphragmatic/phrenic nerve stimulator is a device that provides electrical stimulation of a patient’s phrenic nerve to contract the diaphragm rhythmically and produce breathing in patients who have hypoventilation (a state in which an abnormally low amount of air enters the lungs) caused by brain stem disease, high cervical spinal cord injury, or chronic lung disease. The stimulator consists of an implanted receiver with electrodes that are placed around the patient’s phrenic nerve and an external transmitter for transmitting the stimulating pulses across the patient’s skin to the implanted receiver.

(b)Classification. Class III (premarket approval).

(c)Date premarket approval application (PMA) or notice of completion of a product development protocol (PDP) is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before July 7, 1986 for any implanted diaphragmatic/phrenic nerve stimulator that was in commercial distribution before May 28, 1976, or that has on or before July 7, 1986 been found to be substantially equivalent to an implanted diaphragmatic/phrenic nerve stimulator that was in commercial distribution before May 28, 1976. Any other implanted diaphragmatic/phrenic nerve stimulator shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 51 FR 12101, Apr. 8, 1986]

Sec. 882.5840 Implanted intracerebral/subcortical stimulator for pain relief.

(a)Identification. An implanted intracerebral/subcortical stimulator for pain relief is a device that applies electrical current to subsurface areas of a patient’s brain to treat severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed within a patient’s brain and an external transmitter for transmitting the stimulating pulses across the patient’s skin to the implanted receiver.

(b)Classification. Class III (premarket approval).

(c)Date premarket approval application (PMA) or notice of completion of a product development protocol (PDP) is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 1, 1989, for any implanted intracerebral/subcortical stimulator for pain relief that was in commercial distribution before May 28, 1976, or that has on or before March 1, 1989, been found to be substantially equivalent to an implanted intracerebral/subcortical stimulator for pain relief that was in commercial distribution before May 28, 1976. Any other implanted intracerebral/subcortical stimulator for pain relief shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 53 FR 48621, Dec. 1, 1988]

Sec. 882.5850 Implanted spinal cord stimulator for bladder evacuation.

(a)Identification. An implanted spinal cord stimulator for bladder evacuation is an electrical stimulator used to empty the bladder of a paraplegic patient who has a complete transection of the spinal cord and who is unable to empty his or her bladder by reflex means or by the intermittent use of catheters. The stimulator consists of an implanted receiver with electrodes that are placed on the conus medullaris portion of the patient’s spinal cord and an external transmitter for transmitting the stimulating pulses across the patient’s skin to the implanted receiver.

(b)Classification. Class III (premarket approval).

(c)Date PMA or notice of completion of a PDP is required. A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before December 26, 1996 for any implanted spinal cord stimulator for bladder evacuation that was in commercial distribution before May 28, 1976, or that has, on or before December 26, 1996 been found to be substantially equivalent to an implanted spinal cord stimulator for bladder evacuation that was in commercial distribution before May 28, 1976. Any other implanted spinal cord stimulator for bladder evacuation shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 52 FR 17740, May 11, 1987; 61 FR 50708, Sept. 27, 1996]

Sec. 882.5860 Implanted neuromuscular stimulator.

(a)Identification. An implanted neuromuscular stimulator is a device that provides electrical stimulation to a patient’s peroneal or femoral nerve to cause muscles in the leg to contract, thus improving the gait in a patient with a paralyzed leg. The stimulator consists of an implanted receiver with electrodes that are placed around a patient’s nerve and an external transmitter for transmitting the stimulating pulses across the patient’s skin to the implanted receiver. The external transmitter is activated by a switch in the heel in the patient’s shoe.

(b)Classification. Class III (premarket approval).

(c)Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP for a device described in paragraph (b) of this section is required to be filed with the Food and Drug Administration on or before July 13, 1999 for any implanted neuromuscular stimulator that was in commercial distribution before May 28, 1976, or that has, on or before July 13, 1999, been found to be substantially equivalent to an implanted neuromuscular stimulator that was in commercial distribution before May 28, 1976. Any other implanted neuromuscular stimulator shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 52 FR 17740, May 11, 1987; 64 FR 18329, Apr. 14, 1999]

Sec. 882.5870 Implanted peripheral nerve stimulator for pain relief.

(a)Identification. An implanted peripheral nerve stimulator for pain relief is a device that is used to stimulate electrically a peripheral nerve in a patient to relieve severe intractable pain. The stimulator consists of an inplanted receiver with electrodes that are placed around a peripheral nerve and an external transmitter for transmitting the stimulating pulses across the patient’s skin to the implanted receiver.

(b)Classification. Class II (performance standards).

Sec. 882.5880 Implanted spinal cord stimulator for pain relief.

(a)Identification. An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient’s spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient’s spinal cord and an external transmitter for transmitting the stimulating pulses across the patient’s skin to the implanted receiver.

(b)Classification. Class II (performance standards).

Sec. 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient’s skin to treat pain.

(b)Classification. Class II (performance standards).

Sec. 882.5900 Preformed craniosynostosis strip.

(a)Identification. A preformed craniosynostosis strip is a plastic strip used to cover bone edges of craniectomy sites (sites where the skull has been cut) to prevent the bone from regrowing in patients whose skull sutures are abnormally fused together.

(b)Classification. Class II (performance standards).

Sec. 882.5910 Dura substitute.

(a)Identification. A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).

(b)Classification. Class II (performance standards).

Sec. 882.5940 Electroconvulsive therapy device.

(a)Identification. An electroconvulsive therapy device is a device used for treating severe psychiatric disturbances (e.g., severe depression) by inducing in the patient a major motor seizure by applying a brief intense electrical current to the patient’s head.

(b)Classification. Class III (premarket approval).

(c)Date PMA or notice of completion of a PDP is required. No effective date has been established of the requirement for premarket approval. See 882.3.

[44 FR 51730-51778, Sept. 4, 1979, as amended at 52 FR 17740, May 11, 1987]

Sec. 882.5950 Neurovascular embolization device.

(a)Identification . A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see 870.3300.

(b)Classification . Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see 882.1(e).

[69 FR 77900, Dec. 29, 2004]

Sec. 882.5960 Skull tongs for traction.

(a)Identification. Skull tongs for traction is an instrument used to immobilize a patient with a cervical spine injury (e.g., fracture or dislocation). The device is caliper shaped with tips that penetrate the skin. It is anchored to the skull and has a heavy weight attached to it that maintains, by traction, the patient’s position.

(b)Classification. Class II (performance standards).

Sec. 882.5970 Cranial orthosis.

(a)Identification. A cranial orthosis is a device that is intended for medical purposes to apply pressure to prominent regions of an infant’s cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.

(b)Classification. Class II (special controls) (prescription use in accordance with 801.109 of this chapter, biocompatibility testing, and labeling (contraindications, warnings, precautions, adverse events, instructions for physicians and parents)).

[63 FR 40651, July 30, 1998]

Sec. 882.5975 Human dura mater.

(a)Identification. Human dura mater is human pachymeninx tissue intended to repair defects in human dura mater.

(b)Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Human Dura Mater.” See 882.1(e) for the availability of this guidance.

(c)Scope. The classification set forth in this section is only applicable to human dura mater recovered prior to May 25, 2005.

[68 FR 70436, Dec. 18, 2003, as amended at 76 FR 36993, June 24, 2011]

4 Comments

  1. Greeting from across the sea. precise article I must return for more.

  2. Yeah bookmaking this wasn’t a high risk determination great post!

Trackbacks

  1. ECT bisa membantu menangani Depresi (ECT can help depression) « Hana Medical Consulting
  2. Definisi Alat-Alat Kesehatan (Neurology, FDA) | Electrician Dallas

Leave a comment